TICARE® II Type Automated External Defibrillators (AEDs)
Display
Screen Type: High-resolution display
Screen Size: 8 inch
Languages: 5 to Choose
Sweep Speed: 25mm/sec
Information: Heart Rate, Lead/Pads, Alarm, On/Off, SpO2, AED
Functions and Prompts Alarm Selection and Limits Delivered Energy
5 Languages available: English/Russian/Spanesh/Chinese/Turkish
Defibrillator
Waveform: Biphasic
Charge Time: Less than 7 seconds with a new fully charged battery
Energy Display: Monitor display indicates both selected and delivered energy
Charge Prompt Type: Voice and visual prompts
Electrode Impedance Measurement Range: 0-250 ohms.
ECG Monitoring
Patient Connection: 3-lead ECG cable or 5-lead ECG cable paddles
Lead Selection: Displayed on monitor paddles III lI AVR AVL AVF V
ECG Size: 0.25/0.5/1/1.5/24 cm/mV display on monitor
Heart Rate: 20-300BPM
Heart Rate Alarm: On/Off displayed on monitor user-selectable
Smart Alarms: Beeper/voice prompts indicate shock able rhythm.
Battery
Type: Rechargeable Ni-MH battery 12V
Operating time: For a new fully charged battery: 60 defibrillator discharges at
maximum energy or 3 hours minimum of continuous ECG monitoring. Additional parameters will effect operating time with different functions.
Sp02 Module [Options]
Measurement Range: 30~100%±2% between 80%~90% others±5%
Alarm Range: User set high limit and low limit
Alarm Accuracy:±10% within setting values
Alarm time Accuracy: less than 12 sec
1-Meditech Pulse Oximeter
2-Nellcor Pulse Oximeter
Pacemaker
Type: VVI
Pulse Type: rectangular,constant current
Pulse Amplitude: 0 to 180 mA±10% or 5 mA (whichever is greater).
Pacing Rate: Variable from 30 ppm to 180 ppm±1.5% (increments or decrements by a value of 2 ppm)
Multi-Function Electrode: (MFE) Pads: multipurpose defibrillation/pacing electrodes
Pause: Pacing pulse frequency reduced by a factor of 4 when activated
Refractory Period: NC(WI demand do not have this specification)
Pulse Width: (20+1.5 ms)
Output Protection: Fully defibrillator protected and isolated
Recorder
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.
AED Mode
AED Function: Auto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >= 140bpm and QRS complex wave duration >=1 40ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Manual Mode
Energy Selection: Selectable at 25710 20 30 50 70100 150 200 300 360 joules.
Synchronized Mode: Synchronizes defibrillator pulse to patient's
R-wave. "SYNC" message displayed on monitor.
Recommended Products
1. TICARE® I Type Automated External Defibrillators
2. TICARE® Chest Drainage System 1 Chamber